EPS Bio Technology EasyPlus mini R2N Instrukcja Użytkownika Strona 22

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 47
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 21
21
Troubleshooting Checks Action
9 Was the test strip exposed to open air for a long
period of time?
Repeat the control test with properly stored strips.
9 Was the test strip bottle capped tightly?
This will humidify strips inside. Replace the test
strips.
9 Was the meter functioning well?
You can use the check strip to verify the meter’s
functions.
9 Are the control solutions expired or
contaminated?
Replace with a new control solution to check the
performance of SMBG system.
9 Were test strips and control solutions stored in
cool, dry places?
Repeat the control test with properly stored strips
or control solutions.
9 Did you follow the testing steps properly?
Read Chapter 2 "Control Solution Testing" and test
again. Stop using the meter if you continue to
obtain the inaccurate results. Call Consumer
Services Toll-Free: 1-866-203-2761(Eastern Time,
Mon-Fri 9:00AM~5:00PM).
Przeglądanie stron 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 46 47

Komentarze do niniejszej Instrukcji

Brak uwag